Ultrasonics – Surgical systems – Measurement and declaration of the basic output characteristics
Introduction: Ultrasonic surgical systems, operating in the 20 kHz to 120 kHz range, are used widely in ophthalmology and neurosurgery to fragment or disintegrate and aspirate unwanted tissue. Their commercial use in ophthalmology started in 1970. Their application in neurosurgery followed about 10 years later. Ultrasonic surgical systems are also widely used in oncology surgery. The use of these systems has expanded to areas such as liposuction and wound treatments. This document defines the parameters which characterize the output and performance of open and closed site ultrasonic surgical systems and indicates which parameters should be declared. In addition, measurement procedures are described so that technically qualified people will be able to report on the parameters in a uniform and understandable fashion. An open surgical site is one in which the area of use is large relative to the size of the applicator tip being inserted thus precluding any increase in pressure of the organ due to an imbalance of irrigant flow and suction flow. An example of a closed surgical site is an eye where the incision is closely controlled. This document does not provide any guidance on what is the resultant safety or efficacy of systems described by these parameters. While available data indicate that inertial cavitation is an important component of efficacy for certain applications, other effects such as acoustic streaming can be more important in other applications. Overall, it is important that man
Scope: This document specifies: the essential non-thermal output characteristics of ultrasonic surgical units; NOTE 1 One of the parameters of interest is output acoustic power. This document primarily addresses the low-frequency (under 120 kHz) component of the total delivered energy. The high-frequency component, which relates to cavitation developed at the tip, is discussed in Clause A.4. methods of measurement of these output characteristics; those characteristics to be declared by the manufacturers of such equipment. NOTE 2 In the interest of clarity, a straight tubular shape is used in the basic description of the parameters and measurements to be made. Guidance is provided to the user of this document to adapt the basic methodology described to more complex designs as required. It is recognized that complex designs and vibration patterns are design features of many surgical systems, and therefore it is important that output characteristics be declared for those conditions. This document is applicable to equipment which meets the criteria of a), b) and c) below: a) ultrasonic surgical systems operating in the frequency range 20 kHz to 120 kHz; and b) ultrasonic surgical systems whose use is the fragmentation, emulsification, debridement, or cutting of human tissue, whether or not those effects are delivered in conjunction with tissue removal or coagulation; and c) ultrasonic surgical systems in which an acoustic wave is conducted by means of a specifically designed wave guide to deliver energy to the surgical site. NOTE 3 Examples of these types of systems are surgical aspirators, phacoemulsifiers, intracorporeal lithotripters, end-cutting systems, ultrasonic liposuction systems, etc. NOTE 4 The upper frequency limit has been set to accommodate more recently developed systems operating at higher frequencies than IEC 61847:1998. The techniques of this document are also useful for systems operating at higher frequencies that use the same mechanisms of action. This document is not applicable to: lithotripsy equipment which uses extracorporeally induced pressure pulses, focused through liquid conducting media and the soft tissues of the body; surgical systems used as
Contents:
- FOREWORD
- INTRODUCTION
- Scope