Biotechnology — General requirements for the measurement of ultra-low concentration samples of target nucleic acid sequence
Introduction: When measuring ultra-low concentration of nucleic acid samples, the measurements often exhibit significant variation. This large variation is primarily caused by the Poisson distribution and other contributing factors. In addition, the ultra-low concentration makes the samples highly susceptible to issues such as adsorption and contamination. This document summarizes the key considerations and requirements for handling, measurement including both qualification and quantification analysis, and storage of ultra-low concentration sample of target nucleic acids sequence. Analytical methods based on polymerase chain reaction (PCR) are theoretically capable of detecting the presence of one molecule of nucleic acid in the reaction solution, making it possible to amplify and measure the ultra-low concentration target nucleic acid sequence. These methods are used not only in the field of biotechnology, but also in many other areas such as medicine, food, and the environment. However, it is difficult to stably handle, measure, and store ultra-low concentration target nucleic acid sequences. For example, nucleic acids readily adsorb to polypropylene (PP) materials, which is widely used as a base material for labware, so target nucleic acid sequences stored in such containers are easily removed from the sample/reaction solution during storage or experiment and become undetectable. Similarly, if a sample before PCR and a sample after PCR are handled in the same place for the same target n
Scope: This document specifies requirements and considerations for handling, measuring, and storing samples with ultra-low concentration of target nucleic acid sequences, i.e., concentrations corresponding to copy numbers. This document is applicable to nucleic acid detection methods (qPCR, dPCR, isothermal amplification, and NGS).