Significance and Use
5.1 This test method is one of those required to determine if the presence of an item in the B 0 near field may cause injury to individuals during an MR examination or in the MR environment. Other safety issues which should be addressed include, but may not be limited to: magnetically induced torque (see Test Method F2213 ) and radiofrequency (RF) heating (see Test Method F2182 ). The terms and icons in Practice F2503 should be used to mark the item for safety in the magnetic resonance environment.
5.2 Two methods for measurement of the magnetically induced displacement force are given in this standard: ( 1 ) the deflection on a string method, and ( 2 ) the force gauge method.
5.3 Deflection on a String Method— This method is applicable to items small enough to be suspended by a string and attached to a protractor to measure the deflection angle.
5.4 Force Gauge Method— This method is applicable to items that may be placed on a flat, low friction surface at the isocenter level of the MR equipment bore. This method may be used on items that cannot be suspended by a string.
5.5 Acceptable Magnetic Displacement Force— This standard does not specify an acceptable maximum magnetically induced displacement force that will not harm a patient or other individual. The user of this standard shall specify and provide a rationale for the acceptance criterion used.
5.5.1 For example, as a conservative reference point, if the maximum magnetically induced displacement force for the specified magnetic field conditions is less than the force on the implant due to gravity (i.e., its weight), any risk imposed by the application of the magnetically induced force is no greater than any risk imposed by normal daily activity in the Earth’s gravitational field. For some implants, the acceptable magnetic displacement force can be greater than their weight and may be determined with consideration of the tissue adjacent to the implant or means of fixation of the item.
5.5.2 The acceptable static magnetic field strength ( B 0 ) and spatial field gradient for safety (i.e., labeled conditions of use) shall be calculated based on the acceptable magnetic displacement force using the equations in Appendix X3 .
5.6 This test method alone is not sufficient for determining if an item is safe in the B 0 near field.
Scope
1.1 This test method covers the measurement of the magnetically induced displacement force produced by static magnetic field gradients (spatial field gradient) on medical devices and other items intended for MR conditional use in the B 0 near field (i.e., the space around an MR equipment where the static magnetic field strength ( B 0 ) is >200 mT) of the magnetic resonance environment.
1.2 “Medical devices and other items” will be referred to as “items” for the remainder of this test method.
1.3 This test method does not address other possible safety issues which include, but are not limited to: issues of magnetically induced torque, radiofrequency (RF) heating, induced heating, acoustic noise, interaction among items, and the functionality of the item in the magnetic resonance (MR) environment.
1.4 This test method shall be carried out using a cylindrical whole-body MR equipment with a static magnetic field oriented horizontally and parallel to the MR equipment bore.
1.5 The results can be applicable to whole-body clinical MR equipment (e.g., 1.2 T, 1.5 T, 3.0 T, 7.0 T MR equipment). The results could be applicable to other types of MR equipment including head-only or extremity clinical MR equipment.
1.6 When an item is tested according to this standard in the B 0 near field (ASTM F2052 ), there is no need for further testing in the B 0 far field according to Test Method F3715 .
1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.