Medical electrical equipment - Characteristics of electro-optical X-ray image intensifiers - Part 7: Determination of the modulation transfer function

Introduction: Imaging systems are often being evaluated using subjective measures of performance, such as limiting resolution. These methods do not necessarily describe the performance appropriately in imaging tasks that are relevant to the intended use of the system, and are susceptible to the variability of human observers. Linear shift-invariant imaging systems can be conveniently analysed in terms of their TRANSFER functions. The signal TRANSFER of such systems can be specified by the OPTICAL TRANSFER FUNCTION (OTF), which shows the response of the system to sine waves as a function of their spatial frequency. The MODULATION TRANSFER FUNCTION (MTF), the modulus of the OTF, is sufficient for describing the signal TRANSFER of X-RAY IMAGE INTENSIFIERS. Systems are called shift-invariant when the POINT SPREAD FUNCTION does not vary with position. Note that X-RAY IMAGE INTENSIFIERS generally are shift-invariant over a limited area only: the isoplanatic region. The MODULATION TRANSFER FUNCTION can be determined in several ways (see, for example [1] of annex D): from square wave responses; from the Fourier transform of the LINE SPREAD FUNCTION; from the Henkel transform of the POINT SPREAD FUNCTION; from scanning a slit image with a spatial filter. Any method is acceptable if performed correctly. For the purpose of simplicity, this standard elaborates on two methods: the Fourier transform of the LINE SPREAD FUNCTION, referred to as the LSF method, and the spatial filter method. Accurate deter

Scope: This International Standard applies to ELECTRO-OPTICAL X-RAY IMAGE INTENSIFIERS for medical use, as components of diagnostic X-RAY EQUIPMENT. This International Standard describes a method of determining the MODULATION TRANSFER FUNCTION of X-RAY IMAGE INTENSIFIERS.

Contents:

  • FOREWORD
  • INTRODUCTION
  • 1 Scope
  • 2 Normative references
  • 3 Terminology
  • 3.2 Degree of requirements and reading instructions
  • 4 Requirements
  • 4.1 Test set-up
  • 4.2 X-RAY IMAGE INTENSIFIER - Operating conditions
  • 4.3 Input RADIATION
  • 4.4 TEST DEVICE
  • 4.5 Measurement equipment
  • 5 Determination of the MODULATION TRANSFER FUNCTION
  • 5.2 Measurement
  • 5.3 Corrections
  • 5.4 Determination of the LOW-FREQUENCY DROP
  • 5.5 Overall accuracy of determination
  • 6 Presentation of the MODULATION TRANSFER FUNCTION
  • 7 Statement of compliance