Ultrasonics – Pulse-echo scanners – Low-echo sphere phantoms and method for performance testing of grey-scale medical ultrasound scanners applicable to a broad range of transducer types
Introduction: Ultrasonic pulse-echo scanners are widely used in medical practice to produce images of soft- tissue organs throughout the human body. Most ultrasonic pulse-echo scanners produce real- time images of tissue in a scan plane by sweeping a narrow, pulsed beam of ultrasound through the tissue section of interest and detecting the echoes generated by reflection at tissue boundaries and by scattering within tissues. Many newer scanners transmit broad, overlapping ultrasound beams, and apply software beam-forming to synthesize narrow, pulse-echo beam patterns. Generally, the sweep that generates an image frame is repeated at least 20 times per second, giving rise to the real-time aspect of the displayed image. The axes of the pulsed beams generally lie in a plane that defines the scan plane. Various transducer types are employed to operate in a transmit–receive mode to generate and detect the ultrasonic signals. Linear arrays, in which the beam axes are all parallel to one another, resulting in a rectangular image, consist of a line of hundreds of parallel transducer elements with a subset of adjacent elements producing one pulse at a time. Convex arrays are similar to linear arrays but the element arrangements define part of the surface of a short, right, circular cylinder with the array elements parallel to the axis of the cylinder. The radius of curvature of the cylinder (and therefore the array) can have values between 0,5 cm and 7 cm. The convex array generates a sector image,
Scope: This document, which is a Technical Specification, defines terms and specifies methods for quantifying detailed imaging performance of real-time, ultrasound B-mode scanners. Detail is assessed by imaging phantoms containing small, low-echo spherical targets in a tissue- mimicking background and analysing sphere detectability [3]. Specifications are given for phantom properties. In addition, procedures are described for acquiring images, conducting qualitative analysis of sphere detectability, and carrying out quantitative analysis by detecting sphere locations and computing their contrast-to-noise ratios. With appropriate choices in design, results can be applied, for example: to assess the relative ability of scanner configurations (scanner make and model, scan head and console settings) to delineate the boundary of a tumour or identify specific features of tumours; to choose scanner control settings, such as frequency or the number and location of transmit foci, which maximize spatial resolution; to detect defects in probes causing enhanced sidelobes and spurious echoes. The types of transducers used (see sections 7.6 and 10.7 of [1]) with these scanners include: a) phased arrays, b) linear arrays, c) convex arrays, d) mechanical sector scanners, e) 3-D probes operating in 2-D imaging mode, and f) 3-D probes operating in 3-D imaging mode for a limited number of sets of reconstructed 2-D images. The test methodology is applicable for transducers operating in the 1 MHz to 23 MHz frequency range.
Contents:
- FOREWORD
- INTRODUCTION
- Scope
- 7 MHz frequency range