Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals
Introduction: As noted in the introduction to ISO 11135, when determining the suitability of ethylene oxide (EO) for sterilization of medical devices, it is important to ensure that the levels of residual EO and ethylene chlorohydrin (ECH) pose a minimal risk to the patient in intended product use. Therefore, it is important that the use of alternative materials and sterilization processes are considered during product design and development. EO is known to exhibit a number of biological effects. In the development of this document, consideration was given to these effects, which include irritation, organ damage, mutagenicity, carcinogenicity, and reproductive effects in humans and animals. Similar consideration was given to the harmful effects of ECH and ethylene glycol (EG). ECH can be formed when EO comes into contact with free chloride ions, whereas EG is a hydrolytic reaction product of EO and water. In practice, for most devices, exposure to EO and ECH is considerably lower than the maximum allowable limits established according to this document. No allowable limits are set for EG because risk assessment indicated that when EO residuals are controlled, it is unlikely that biologically significant residuals of EG would be present. Requirements herein are in addition to the biological evaluation requirements as indicated in ISO 10993-1. The biological evaluation, combined with the EO-sterilization process residual limits, form the justification that an EO-sterilized device is safe for
Scope: This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K. EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228]. NOTE This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.