Sterilization of health care products — Chemical indicators — Part 6: Type 2 indicators and process challenge devices for use in performance testing of small steam sterilizers
Introduction: This document includes a description of both hollow and porous process challenge devices (PCDs) and their performance requirements, along with methods by which an alternative PCD can be shown to have equivalent performance to that of the reference PCD. Small sterilizers unable to accommodate a sterilization module [rectangular parallelepiped of dimensions 300 mm (height) × 600 mm (length) × 300 mm (width)] cannot be tested using the tests described in EN 285 for large sterilizers for wrapped goods and porous loads because — the chamber size of a small steam sterilizer according to EN 13060 is unable to accommodate the standard test pack from EN 285, and — the efficacy of the tests is impaired when the test pack occupies a large proportion of the chamber volume (>20 % chamber volume). Indicators described in this document are intended to be used in conjunction with appropriate PCDs to show penetration of steam into the PCD. The reference indicator systems and alternative indicator systems pose specified challenges to air removal and steam penetration. The devices described in this document are intended for use only in small steam sterilizers conforming to EN 13060 to monitor steam penetration in type B cycles and some type S cycles. NOTE Even though the hollow load was originally designed as a type test in EN 867-5 (withdrawn standard replaced by this document) to test the performance of small steam sterilizers conforming with EN 13060, the same test is also used in other stan
Scope: This document specifies the performance requirements and test methods for hollow devices and porous devices as well as the chemical indicators and biological indicators that are utilized within these devices for testing a specific steam penetration performance of type B cycles and some type S cycles of small steam sterilizers according to EN 13060. NOTE The hollow and porous devices described in this document are not intended for use as surrogate devices for hollow and porous medical devices used in health care facilities. a) Chemical indicators used with a porous device specified in this document are designed to demonstrate the adequacy of steam penetration into a porous device in small steam sterilizers (see EN 13060). This document specifies the requirements for: — a reference porous device (RPD) as a reference device by which alternative porous indicator systems (APISs) can be shown to be equivalent in performance according to this document, i.e. a textile test pack in which steam penetration is judged by thermometric means; — an alternative porous chemical indicator system equivalent in performance to the RPD, i.e. an APIS, usually commercially manufactured, of any design. b) Chemical indicators used with a hollow load device specified in this document are designed to demonstrate the adequacy of steam penetration into a narrow lumen (previously known as hollow load A) in small steam sterilizers (see EN 13060). This document specifies the requirements for: — a reference hollow device (RHD) used as a reference device in this document, i.e. a lumened device with attached capsule in which steam penetration is judged by inactivation or survival of a specified biological indicator; — an alternative hollow device: — employing the same specific test load as defined for the RHD and a chemical indicator designed specifically for use in the reference hollow test load, i.e. a lumened device with an attached capsule in which steam penetration is judged by visual examination of a chemical indicator; — equivalent in performance to the RHD, i.e. an alternative hollow device, usually commercially manufactured, of any design. © ISO 2022 – All rights reserved ISO 11140-6:2022