Water quality — Radium-226 — Part 3: Test method using coprecipitation and gamma-ray spectrometry
Introduction: Radionuclides are present throughout the environment; thus, water bodies (e.g. surface waters, ground waters, sea waters) contain radionuclides, which can be of either natural or anthropogenic origin. — Naturally-occurring radionuclides, including 3H, 14C, 40K, and those originating from the thorium and uranium decay series (see Annex A), in particular 210Pb, 210Po, 222Rn, 226Ra, 228Ra, 227Ac, 231Pa, 234U, and 238U, can be found in water bodies due to either natural processes (e.g. desorption from the soil, runoff by rain water) or released from technological processes involving naturally occurring radioactive materials (e.g. mining, mineral processing, oil, gas and coal production, water treatment, and the production and use of phosphate fertilisers). — Anthropogenic radionuclides, such as 55Fe, 59Ni, 63Ni, 90Sr, 99Tc, transuranic elements (e.g. Np, Pu, Am, Cm) and some gamma emitting radionuclides, such as 60Co, 137Cs, can also be found in natural waters. Small quantities of anthropogenic radionuclides can be discharged from nuclear facilities to the environment as a result of authorized routine releases. The radionuclides present in liquid effluents are usually controlled before being discharged into the environment[1] and water bodies. Anthropogenic radionuclides used in medical and industrial applications can be released to the environment after use. Anthropogenic radionuclides are also found in waters due to contamination from fallout resulting from above-ground nuclear
Scope: This document specifies a method to determine radium-226 (226Ra) activity concentration in all types of water by coprecipitation followed by gamma-ray spectrometry (see ISO 20042[7]). The method covers the measurement of soluble 226Ra activity concentrations greater than 0,002 Bq·l−1 using a sample volume of up to 100 l of any water type. For water samples with a volume of less than a volume of 1 l, direct gamma-ray spectrometry can be performed following ISO 10703 but with a higher detection limit. The typical detection limit for samples of 1 l to 5 l is in the range of 0,002 to 0,000 40 Bq·l−1[8]. NOTE This test method can be adapted to determine other naturally occurring isotopes of radium, such as 223Ra, 224Ra and 228Ra, if the respective ingrowth periods are taken into account.