Washer-disinfectors — Part 6: Requirements and tests for washer-disinfectors employing thermal disinfection for non-critical medical devices and health care equipment
Introduction: This document is the sixth part of the ISO 15883 series of standards specifying the general requirements and performance of washer-disinfectors (WD). The requirements given in this document apply to WD used for cleaning and thermal disinfection of non-critical medical devices (i.e. not penetrating skin or contacting mucosal membranes) and for other devices or equipment for use without further treatment in health care settings. Such reusable equipment is cleaned and disinfected, but it is inappropriate or impractical to process it in a WD intended for surgical instruments (see ISO 15883-2), for human waste containers (see ISO 15883-3), for endoscopes (see ISO 15883-4), or for non-critical thermolabile medical devices and health care equipment (see ISO 15883-7). Examples of non-critical medical devices and health care equipment include: — laryngoscope handles; — washbowls and bowls; — cleaning equipment (buckets); — footwear; — transport carriers, trolleys and transport carts; — bed frames, wheelchairs, aids for the disabled. Fields of application within the scope of the ISO 15883 series can include laboratory, veterinary and dental use, and other specific applications such as washer-disinfectors for the disinfection of crockery and cutlery intended for use with immunologically compromised patients. Requirements for washer-disinfectors for other applications are specified in other parts of ISO 15883. The efficacy of disinfection can be impaired if soil removal is incomplete bef
Scope: This document specifies particular requirements for washer-disinfectors (WD) intended for use when the level of assurance of disinfection that is necessary can be achieved by cleaning and thermal disinfection (A0 not less than 60) and does not require an independent automated record of critical processes to be kept. It is intended to be used in conjunction with ISO 15883-1, which specifies general requirements for WD. The range of products on which WD of this particular type can be used is restricted to non-invasive and non-critical devices and equipment (i.e. not penetrating skin or contacting mucosal surfaces). NOTE Thermal disinfection can be achieved by rinsing the load with hot water, exposure to steam, or combination of the two. This document does not cover powered devices, lumened devices, and other semi-critical and critical medical devices. Devices identified within the scopes of ISO 15883-2, ISO 15883-3, ISO 15883-4, and ISO 15883-7 do not fall within the scope of this document.