Surface chemical analysis — Secondary ion mass spectrometry — Linearity of intensity scale in single ion counting time-of-flight mass analysers
Introduction: For the quantitative analysis of materials using secondary ion mass spectrometry (SIMS), measurements are made of the spectral intensities. Nonlinearity in the instrument intensity scale, unless corrected, leads directly to errors in the relative amounts of material determined at surfaces and in-depth profiles. In general, intensity scales are linear at very low count rates, or more correctly low counts per pulse, but become progressively nonlinear as the count rates rise. Measurements of intensity rely on the measurement system delivering an intensity signal fixed in proportion to the intensity being measured. In counting systems, this proportionality is expected to be unity. If this proportionality varies with the signal level or counting rate, the measurement system is said to be nonlinear. It is rare for nonlinearities below 1 % to be treated as significant. The intensity scale nonlinearity can exceed 1 % for count rates that exceed 5 % of the maximum permissible count rate[2]. For many instruments, the nonlinearity behaviour will not vary significantly from month-to-month, provided the detection system is correctly set. For these instruments, the count rate can be corrected, using the relevant relationship, so that the corrected intensity is then linear for a greatly extended fraction of the maximum obtainable count rate. This correction to the intensity scale might or might not already be available in the instrument's data capture or processing computer. In this documen
Scope: This document specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in single ion counting time-of-flight (TOF) secondary ion mass spectrometers using a test based on isotopic ratios in spectra from poly(tetrafluoroethylene) (PTFE). It also includes a method to correct for intensity nonlinearity arising from intensity lost from a microchannel plate (MCP) or scintillator and photomultiplier followed by a time-to-digital converter (TDC) detection system caused by secondary ions arriving during its dead-time. The correction can increase the intensity range for 95 % linearity by a factor of up to more than 50 so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction formulae have been shown to be valid.