Medical devices - Connectors for reservoir delivery systems for healthcare applications - Part 8: Citrate-based anticoagulant solution for apheresis applications
Introduction: During the development of the series of standards for small-bore connectors for liquids and gases in healthcare applications (see ISO 80369) it became clear that the risk of misconnections was not limited to the patient access connectors and that the whole reservoir system needed to be considered. The possible misconnection between apheresis AC reservoir connectors and spikes was also reviewed. However, as apheresis AC reservoir connectors are not exactly within the definition of small- bore connectors it was decided to develop this separate standard for these connectors, taking into account the risks of misconnections with other devices such as IV bags, and blood pressure cuffs that are likely to be used on the same patient. This document, therefore, defines the apheresis AC reservoir connectors in geometry, material, and performance to permit them being analyzed (and avoided) by other connectors intended for non‐ apheresis applications (i.e. the other application parts of the ISO 18250 series). This document is not a device standard as it specifies only the dimensions of the interfaces for the apheresis AC reservoir connectors used on the disposable tubing sets and solution reservoirs. vi © ISO 2018 – All rights reserved
Scope: This document specifies dimensions and requirements for the design and functional performance of apheresis anticoagulant (AC) reservoir connectors. This document does not specify the dimensions and requirements for the medical devices or accessories that use these connectors. Such requirements are given in particular International Standards for specific medical devices or accessories. The following examples of medical devices are intended to use the connectors of this document: — Apheresis tubing sets and mating reservoirs containing citrate-based anticoagulant solution. — Apheresis tubing sets may include, but are not limited to those for use in blood collection, therapeutic applications, and plasma collection. Some apheresis sets are manufactured with a pre-connected reservoir containing citrate-based anticoagulant solution. These medical devices are not intended to use the connectors of this document. NOTE Manufacturers are encouraged to incorporate the connectors specified in this document into apheresis devices or accessories, even if not currently required by the particular medical device standards. It is expected that when the particular medical device standards are revised, requirements for apheresis AC reservoir connectors, will be included.