Sterilization of health care products - Biological and chemical indicators - Test equipment
Introduction: To test the performance of biological and chemical indicators, specific test equipment is required. This document specifies the performance requirements for the test equipment to be used to establish the response of biological and chemical indicators to critical process variables. This document does not apply to test equipment for indicators used in irradiation, isolator/room biodecontamination (at atmospheric pressure), or low temperature steam and formaldehyde processes. Resistometers constitute test equipment designed to create precise and repeatable sterilizing environments, allowing the evaluation of their effect on biological inactivation kinetics, chemical reactions, material degradation and product bioburden. Resistometers allow precise variation of the environmental conditions and cycle sequences in order to produce controlled physical studies. When used with the defined test methods given in the appropriate parts of ISO 11138 for biological indicators and ISO 11140 for chemical indicators, the results of these studies can be used to demonstrate conformance of biological indicators and chemical indicators to these standards. Resistometers differ from conventional sterilizers. Instrumentation selection and control requirements for resistometers are based upon mathematical models in which rates of reaction, measurement accuracy and process control requirements are evaluated to quantify the effects induced by test equipment-controlled variables. The requirements for acc
Scope: This document specifies the requirements for test equipment to be used to: — test biological indicators for steam, ethylene oxide gas and dry heat sterilization processes for conformity to the requirements given in ISO 11138 series; — test chemical indicators for steam, ethylene oxide gas, dry heat and vaporized hydrogen peroxide sterilization processes for conformity to the requirements given in ISO 11140-1:2014. This document also provides informative methods useful in characterizing the performance of biological and chemical indicators for intended use and for routine quality control testing. This document does not specify requirements for test equipment for processes specifically for testing chemical and biological indicators intended to monitor isolator and room biodecontamination processes at atmospheric pressure. ISO 11138-2:2017, ISO 11138-3:2017, ISO 11138-4:2017 and ISO 11140-1:2014 require the use of resistometers specified in this document, and these resistometers are used in conjunction with the test methods specified in the appropriate parts of ISO 11138 series and ISO 11140 series. Resistometers for low temperature steam and formaldehyde indicators are not included in this document. Test methods using laboratory apparatus for low temperature steam and formaldehyde are included in ISO 11138-5:2017. Test equipment for testing Type 2 (e.g. Bowie Dick) chemical indicators are specified in ISO 11140-3:2007, ISO 11140-4:2007, and ISO 11140-5:2007.