Surface chemical analysis - Determination of lateral resolution and sharpness in beam based methods with a range from nanometres to micrometres
Introduction: Secondary ion mass spectrometry (SIMS), Auger electron spectroscopy (AES) and X-ray photoelectron spectroscopy (XPS) are surface-analytical techniques that are used to generate chemical maps of surfaces and line scans across surfaces. These techniques can have lateral resolutions for instance as good as 10 nm for AES, 50 nm for SIMS and 5 µm for laboratory XPS and can cover areas as large as many square millimetres by using stitching techniques. Different instruments generate images with different lateral resolutions. Moreover, an analyst needs to have a suitable method to measure the lateral resolution of an instrument for any given settings. In this way, analysts can obtain the optimum lateral resolution from a given instrument, appropriate to the analytical requirements, in a consistent and clear way. The ability of the analyst to realize these resolutions in an effective way will, of course, also depend on the quality of the signal levels obtained and the level of noise. Resolution is a quality parameter of images and line scans and describes the performance of imaging instruments used to deliver them. This document is based on ISO/TR 19319:2013, which explains theoretical backgrounds of a determination of resolution and sharpness parameters used to express the performance of imaging instruments[1]. This document describes different methods for the determination of lateral resolution in beam-based methods as AES, SIMS and XPS. These are (a) the straight edge method, (b) t
Scope: This document describes methods for measuring lateral resolution and sharpness in imaging surface chemical analysis. It applies to all methods of surface analysis which use a beam to analyse the chemical composition of surfaces under defined settings of an instrument. It applies to scanning instruments, where a finely focused beam is scanned over the sample in a preselected field of view, as well as to full field imaging instruments, where the field of view is simultaneously imaged by a broad beam, an imaging lens system and a pixelated detector. The methods for measuring lateral resolution and sharpness are — the straight edge method; — the narrow line method; — the grating method. This document applies to instruments and methods that provide information on layers with nanometre thicknesses and to surfaces with nanometre‐sized structures and individual nano‐objects.