Surface chemical analysis - Electron spectroscopies - Procedures for identifying, estimating and correcting for unintended degradation by X-rays in a material undergoing analysis by X-ray photoelectron spectroscopy
Introduction: The basis of X-ray photoelectron spectroscopy is irradiation of a sample surface by soft X-rays and examination of the excited emission in the form of photo-electrons and Auger electrons. In its most widely used mode, the X-ray flux is of low intensity and spread over a large area. Thus, the technique is generally regarded as one of the least destructive of the available “beam” techniques used for the surface chemical analysis of materials. However, since the time of its inception as a technique for surface analysis, there have been reports of changes in composition arising during the course of analysis.[1]- [4] These reports indicated that, for some materials, a form of degradation during analysis needs to be taken into account and, where possible, a correction made. This International Standard addresses these issues and describes a method by which the extent of degradation can be estimated and a suitable correction obtained. © ISO 2016 – All rights reserved
Scope: This International Standard provides a simple procedure for identifying, estimating and correcting for unintended degradation in the elemental composition or chemical state of a material which occurs as a result of X-radiation during the time that a specimen material is exposed to the X-rays used in X-ray photoelectron spectroscopy (XPS). This International Standard does not address comparisons between different types of material nor does it address the mechanisms, depth, or chemical nature of the degradation that occurs. The correction procedure proposed is only valid if the changes are caused by the X-rays and result in less than a 30 % reduction or increase in intensity of a chosen photoelectron peak from the sample material. 2 Terms and definitions For the purposes of this document, the following terms and definitions apply. 2.1 region part of the photo-excited spectrum chosen for detailed acquisition and analysis Note 1 to entry: The region may be chosen because it contains a major or minor peak of a given element or to represent the shape or slope of a background within that energy range, e.g. a detail scan. Note 2 to entry: This usage of region is not to be confused with the area of analysis. 2.2 time zero time at which the X-rays start to irradiate the sample 3 Symbols and abbreviated terms AZ deduced linear rate of change of It,Z as a result of degradation for a given element or state atomic fraction of contamination carbon from the quantification computation dcontamina- tion thickness of a contamination layer on the surface of the sample degradation index kinetic energy, in eV, of the detected electrons FWHM full width at half maximum (intensity) IZ,corrected intensity of a given photoelectron peak after correction for the formation of a layer of con- tamination IZ,measured measured intensity of a given photoelectron peak that is influenced by the presence of a layer of contamination I0,Z undegraded photoelectron intensity of a given element or state, Z If,S final photoelectron intensity of a given element in the survey spectrum Ii,S initial photoelectron intensity of a given element in the survey spectrum INTERNATIONAL STANDARD ISO 18554:2016(E) © IS