Sterilization of health care products — Biological indicators — Part 8: Method for validation of a reduced incubation time for a biological indicator
Introduction: A biological indicator incubation time is the minimum period of cultivation required before making a final determination that a biological indicator is negative (shows no growth). The reference incubation time for biological indicators for established sterilization processes such as moist heat and ethylene oxide is 7 d (see ISO 11138-1:2017). In some instances where biological indicator results are needed as part of the product release process, a 7-day incubation time might not be practical. This is especially the case where biological indicators are used to monitor sterilization processes in hospitals or other health care facilities such as dental or general practitioner offices. The purpose of a reduced incubation time procedure is to demonstrate recovery of the surviving test organisms within the specified reduced incubation time period. The reduced incubation time is a function of the test method and conditions used to establish the incubation time and is independent of the process parameters for the sterilization method used to deliver the lethality. Biological indicators with an incubation time of less than 7 d (a Reduced Incubation Time, or RIT) have been in use since the 1970s. The methodology to determine the RIT was originally created by the biological indicator manufacturers. Later, the United States Food and Drug Administration published guidance for manufacturers seeking regulatory clearance to market biological indicators to health care facilities in the United
Scope: 1.1 This document specifies the requirements for a test method to be utilized to establish or confirm a reduced incubation time (RIT) that is shorter than the 7-day reference incubation time specified in 7.3.2 of ISO 11138-1:2017 for biological indicators used to monitor moist heat sterilization processes or ethylene oxide (EO) sterilization processes. NOTE For biological indicators used for EO sterilization, the stated RIT is applicable to 100 % EO processes or processes that use EO blends, regardless of the product load. 1.2 This document is applicable to manufacturers of biological indicators (BIs) and to end users of BIs who intend to, if required by their quality system, establish, validate or confirm a RIT. 1.3 This document does not apply to biological indicators used to monitor dry heat, low temperature steam formaldehyde (LTSF) or vaporized hydrogen peroxide (VH2O2) sterilization processes. NOTE The method described in this document to establish a RIT for biological indicators used to monitor moist heat or EO sterilization processes has been used extensively for many years. However, there is limited experience in use of this method to establish a RIT for biological indicators used to monitor dry heat, low temperature steam formaldehyde or vaporized hydrogen peroxide sterilization processes. This document, therefore, does not include these sterilization processes.