Washer-disinfectors — Part 3: Requirements and tests for washer-disinfectors employing thermal disinfection for human waste containers
Introduction: This document is the third part in the ISO 15883 series of standards specifying the performance of washer- disinfectors (WD) and the general requirements for performance applicable to WD for human waste containers. The requirements given in this document apply to WD used for emptying, flushing, washing, and thermally disinfecting human waste containers intended for reuse such as: — portable sanitary pans; — supports for single-use bedpans; — hospital bowls for containment of human waste; — urine collection containers; — suction bottles; — products similar to the above and used for similar purposes. Fields of application within the scope of the ISO 15883 series of standards can include laboratory, veterinary, dental and pharmaceutical applications, and other specific applications, such as WD for bedsteads and transport carts and the disinfection of crockery and cutlery intended for use with immunologically compromised patients. Requirements for WD for other applications are specified in other parts of the ISO 15883 series of standards. In order to reduce the risk of spillage and the generation of aerosols, most machines incorporate means to empty human waste containers automatically, e.g. by the action of closing the WD door. Where equipment does not provide automatic emptying facilities, extra care is needed by the user to avoid exposure to human waste and contamination of the work environment including the generation of aerosols. The reliability of a human waste container WD
Scope: This document specifies requirements for washer-disinfectors (WD) that are intended to be used for emptying, flushing, washing and thermal disinfection of non-critical devices in the form of human waste containers by one operating cycle. This document is intended to be used in conjunction with the general requirements specified in ISO 15883-1:2024, except for those specified in 4.1.1, and with the requirements of ISO 15883-5:2021, except for those specified in 4.1.2.