Methods for the calibration of vibration and shock transducers - Part 44: Calibration of field vibration calibrators
Introduction: ISO 16063 comprises a series of documents dealing with methods for the calibration of vibration and shock transducers. This document focuses on field vibration calibrators (FVCs). In this context, FVCs are mainly used for in situ checks of vibration and shock transducers providing sine wave vibration at known frequency and magnitude under field conditions. The FVC acts as a calibrated vibration source for in situ checks of transducer sensitivity that is specified or requested, for example, in ISO 8041-1 or in ISO 8042. Operational frequency and acceleration RMS value of FVCs are usually 160 Hz or 159,2 Hz, and 3,16 m/ s2, 9,81 m/s2 or 10 m/s2, respectively, which are widely accepted as reference conditions. There are also FVCs with selectable acceleration magnitudes and frequencies. In comparison with stationary calibration systems, FVCs have limitations in shaker power and inertial mass. Therefore, they can be unsuitable for heavy test objects, high acceleration magnitudes and large displacements. Using the calibration procedure described by this document, the acceleration generated by an FVC can be traceable, through chain of calibration, to a primary or national standard as defined by ISO/ IEC Guide 99 (“the VIM”) with associated uncertainty defined by ISO/IEC Guide 98-3 (“the GUM”). © ISO 2018 – All rights reserved
Scope: This document specifies the instrumentation and procedure to be used for performing calibration of field vibration calibrators (FVCs). It is not applicable to FVCs used for the calibration of transducers. These are covered by ISO 16063-21. Procedures and requirements of in situ calibration by FVC are beyond the scope of this document. Annex B provides more information on the application of FVC.