Systems for evacuation of plume generated by medical devices
Introduction: Certain surgical, diagnostic, and therapeutic techniques can generate noxious airborne contaminants (plume) as by-products, particularly from procedures that include the cutting, ablation, cauterization, or mechanical manipulation of target tissue by energy-based devices such as lasers, electrosurgery generators, broadband light sources, and ultrasonic instruments. Energy-based contact with articles such as tubing, swabs, and skin preparation solutions can produce additional chemicals. This document was developed in response to awareness of the potential hazards to patients and staff of plume generated by these techniques in healthcare settings. Plume can contain a variety of contaminants: airborne chemicals, particulates, ultrafine particles, aerosols, gases, vapours, volatile organic compounds, tissue fragments, cellular material and blood-borne pathogens, posing a hazard to exposed persons. Additionally, plume reduces the clinician’s ability to clearly see the operative field, resulting in unsafe operating conditions. This document specifies requirements for systems for evacuation of plume generated in healthcare facilities. It is intended for those persons involved in the design, construction, inspection, and operation of healthcare facilities. Those persons involved in the design, manufacture, installation, testing, and use of equipment and components for plume evacuation systems should also be aware of the contents of this document. This document provides the informatio
Scope: 1.1 This document specifies requirements and guidelines for systems and equipment used to evacuate plume generated by medical devices. 1.2 This document applies to all types of plume evacuation systems (PESs), including a) portable; b) mobile; c) stationary, including dedicated central pipelines; d) PESs integrated into other equipment; e) PESs for endoscopic procedures (e.g., minimally invasive, laparoscopic). 1.3 This document applies to all healthcare facilities where PESs are used, including, but not limited to a) surgical facilities; b) medical offices; c) cosmetic treatment facilities; d) medical teaching facilities; e) dental clinics; f) veterinary facilities. 1.4 This document provides guidance on the following aspects of PESs: a) importance; b) purchasing; c) design; d) manufacture; e) documentation; f) function; g) performance; h) installation; i) commissioning; j) testing; k) training; © ISO 2024 – All rights reserved ISO 16571:2024(en) l) use; m) risk assessment; n) servicing; o) maintenance. 1.5 This document does not apply to the following: a) anaesthetic gas scavenging systems (AGSSs) which are covered in ISO 7396-2; b) medical vacuum systems which are covered in ISO 7396-1; c) heating, ventilation, and air-conditioning (HVAC) systems; d) aspects of laser safety other than airborne contamination; and e) aspects of electrosurgery, electrocautery, and mechanical surgical tools other than airborne contamination produced by such equipment resulting from interaction with tissue or materials. 2 Normative references The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 3744:—1), Acoustics — Determination of sound power levels and sound energy levels of noise sources using sound pressure — Engineering methods for an essentially free field over a reflecting plane ISO 7000, Graphical symbols for use on equipment — Registered symbols ISO 7396-1, Medical gas pipeline systems — Part 1: Pipeline systems for compr