SAME AS IEC 61010-2-011 Equipment included in scope This Part 2 of IEC 61010 specifies particular safety requirements for the following types a) to c) of electrical equipment and their accessories, wherever they are intended to be used, whenever that equipment incorporates REFRIGERATING SYSTEMS as an integral part of, or separate from, the equipment and the equipment is in direct control of the REFRIGERATING SYSTEM. This document details all the requirements when up to 150 g of FLAMMABLE REFRIGERANT are used per stage of a REFRIGERATING SYSTEM. Additional requirements beyond the current scope of this document apply if a REFRIGERANT charge of FLAMMABLE REFRIGERANT exceeds this amount. a) Electrical test and measurement equipment This is equipment which by electromagnetic means tests, measures, indicates or records one or more electrical or physical quantities, also non-measuring equipment such as signal generators, measurement standards, power supplies for laboratory use, transducers, transmitters, etc. NOTE 1 This includes bench-top power supplies intended to aid a testing or measuring operation on another piece of equipment. Power supplies intended to power equipment are within the scope of IEC 61558 (see 1.1.2 h)). This standard also applies to test equipment integrated into manufacturing processes and intended for testing manufactured devices. NOTE 2 Manufacturing test equipment is likely to be installed adjacent to and interconnected with industrial machinery in this application. b) Electrical industrial process-control equipment This is equipment which controls one or more output quantities to specific values, with each value determined by manual setting, by local or remote programming, or by one or more input variables. c) Electrical laboratory equipment This is equipment which measures, indicates, monitors, inspects or analyses materials, or is used to prepare materials, and includes in vitro diagnostic (IVD) equipment. This equipment may also be used in areas other than laboratories; examples include self-test IVD equipment to be used in the home and inspection equipment to be used to check people or material during transportation. For REFRIGERATION SYSTEM with more than 150 g per stage, the requirement of UL 60335-2-89 shall be used. Transcritical REFRIGERATION SYSTEMS shall be evaluated to the requirements of UL 60335-2-89. This document also details the requirements for STIRLING REFRIGERATION SYSTEMS. NOTE 101 Examples for REFRIGERATING EQUIPMENT include, but are not limited to, laboratory equipment such as laboratory refrigerators, freezers, refrigerated display cabinets. It is possible that all or part of the equipment falls within the scope of one or more other Part 2 standards of IEC 61010 as well as within the scope of this standard. In that case, the requirements of those other Part 2 standards will also apply. In particular, if equipment is intended to be used as a centrifuge, the requirements of IEC 61010-2-020 apply. However, when the equipment incorporates a refrigerating system and a heating function where the combination of the two introduces additional or more severe HAZARDS than if treated separately, then it is possible that IEC 61010-2-012 is applicable instead of this Part 2-011. See further information in the flow chart for the selection process and guidance in the Introduction. Equipment excluded from scope This standard does not apply to equipment within the scope of: a) IEC 60065 (Audio, video and similar electronic apparatus); b) IEC 60204 (Safety of machinery – Electrical equipment of machines); c) IEC 60335 (Household and similar electrical appliances); d) IEC 60364 (Electrical installations of buildings); e) IEC 60439 (Low-voltage switchgear and controlgear assemblies); f) IEC 60601 (Medical electrical equipment); g) IEC 60950 (Information technology equipment including electrical business equipment, except as specified in 1.1.3); h) IEC 61558 (Power transformers, power supply units and similar); i) IEC 61010-031 (Hand-held probe assemblies); j) IEC 61243-3 (Live working – Voltage detectors – Part 3: Two-pole low-voltage type). aa) a system that uses ammonia (NH3) as the REFRIGERANT. or equipment incorporating: Computing equipment This standard applies only to computers, processors, etc. which form part of equipment within the scope of this standard or are designed for use exclusively with the equipment. NOTE Computing devices and similar equipment within the scope of IEC 60950 and conforming to its requirements are considered to be suitable for use with equipment within the scope of this standard. However, some of the requirements of IEC 60950for resistance to moisture and liquids are less stringent than those in this standard (see 5.4.4 second paragraph)). This standard applies to equipment: a) To be employed in accordance with ANSI/NFPA 70, National Electrical Code ® (NEC); b) Designed to comply with the general requirements of CAN/CSA C22.2 No. 0 and to be installed in accordance with the Canadian Electrical Code (CEC), Part I, CSA C22.1; or c) Both (a) and (b).